Glow Discharge Optical Emission Spectrometry (GD-OES) - GD-PROFILER 2 by HORIBA Jobin Yvon
Contact | Ing. Eva Švábenská, Ph.D. |
RF Glow Discharge Optical Emission Spectrometry (RF-GD-OES) can provide both the surface and bulk composition to all elements for almost all solid materials, including metals, metal coatings, semiconductors, polymer coatings, glass, ceramics, etc.
RF Glow Discharge Optical Emission Spectrometry (RF-GD-OES) can provide both the surface and bulk composition to all elements for almost all solid materials, including metals, metal coatings, semiconductors, polymer coatings, glass, ceramics, etc.
Polychromator- Paschen-Runge Mounting - 0.5m focal length
Spectral Range: 110nm – 620nm. This spectral range allows the analysis of gas elements (H, O, N, C) at their sensitive wavelengths in the VUV range
Monochromator - The monochromator can measure simultaneously any “n+1” element within a depth profile analysis and any sequence of elements in a bulk program.
Calibrated methods for metallic materials - steel, metal alloys based on Ni, Ti, Al and profile analysis.